Product Data Sheet    hbl 25603

HBL25603



Mobile X-Ray Receptacle

Hubbell Wiring Device-Kellems Product Specifications
 Hospital Product Type  
 Mobile X-Ray Device  Receptacle
 Amperage  60A
 Voltage  250V
 X-Ray Device Type  Cast Aluminum Plate
 Color  
 Description  Hospital Use Only Mobile 60 Amp X-Ray Receptacle w/ Stainless Steel Assembly
 Type  2 Pole, 3 Wire, Grounding
 Certification  Listed to UL Standard 498. Certified to CSA 22.2, No. 42 where applicable.
 IP Suitability  IP30
 Power Contacts Material  .032" (.8) Bronze
 Ground Contacts Material  .032" (.8) Bronze
 Terminal Screw Material  Steel
 Ground Screw Material  Brass
 Mounting Screws Material  steel
 Dielectric Voltage  Withstands 2,000V minimum.
 Current Interrupting  Certified for current interrupting at full rated current.
 Temperature Rise  Max 30 deg C temperature rise at full rated current after 50 cycles of overload at 150% of rated current with direct current
 Terminal Identification  Terminals identified in accordance with UL498 (Brass, White, Green).
 Terminal Accommodation  #4-#8 AWG Copper Conductor only.
 Product Identification  Ratings are a permanent part of the device.
 Flammability  UL 94 V0 and UL 94-5 VA.
 Operating Temperatures  Maximum continuous 75°C, minimum -40°C (w/o impact).




 

Hubbell Wiring Device-Kellems